eurOtox I2P

Pomiarowy Detektor Gazów „eurOtox I2P”
The "eurOtox I2P" Measuring Gas Detector is used to measure the concentration of toxic gases and oxygen. It enables the concentration of two gases to be measured simultaneously, using an internal sensor and an external detector with a 4-20 mA current output. It is a stand-alone device which is a combination of a detector and a control unit.
Mains supply ~230 V, maximum power consumption 15W.
Type of internal sensor (versions):
- O2 oxygen: electrochemical, selective, measuring range : 0-25 % V/V
- carbon monoxide CO: electrochemical, selective, measuring range : 0-500 ppm
- carbon dioxide CO2: infrared, selective, measuring range : 0-5 % V/V
- chlorine Cl2: electrochemical, selective, measuring range : 0-20 ppm
- Sulphur hexafluoride SF6: infrared, selective, measuring range : 0-2000 ppm
External detector connection
- external detector power supply : 12 V DC, max. 12 V DC, max. current consumption 250 mA
- Current input : (2) 4-20 mA , Rwe = 200 Ω
Alarm thresholds : 2 or 3, programmable alarm concentration levels (separately for internal and external detector).
Output type:
- relay (NO) : 3 outputs with programmable functions, max. load 3A/~230V
- Voltage : 2 outputs (WY-AL1, WY-AL2) with programmable functions, max. load 0.5A (total)
- RS485 (Modbus RTU protocol).
Detector functions:
- LCD display (visualisation of detector states, output states, configuration)
- 4-button keypad
- real-time clock
- event memory (1000 events)
- freely configurable inputs and outputs
- user-programmable control algorithm
- predefined configurations (factory settings)
- access protected by passwords.
Device versions:
- eurOtox O2 I2p/E/D/S/RS485 - O2 measurement detector (electrochemical sensor), relay outputs, RS485, 4-20mA output,
possibility to connect an external detector (4-20mA input)
- eurOtox CO I2p/E/D/S/RS485 - CO measurement detector (electrochemical sensor), relay outputs, RS485, 4-20mA output,
Possibility of connecting an external detector (4-20mA input)
- eurOtox CO2 I2p/E/D/S/RS485 - CO2 measuring detector (infrared sensor), relay outputs, RS485, 4-20mA output,
possibility to connect an external detector (4-20mA input)
- eurOtox Cl2 I2p/E/D/S/RS485 - Cl2 measuring detector (electrochemical sensor), relay outputs, RS485, 4-20mA output,
possibility to connect an external detector (4-20mA input)
- eurOtox SF6 I2p/E/D/S/RS485 - SF6 measuring sensor (infrared sensor), relay outputs, RS485, 4-20mA output,
possibility to connect an external detector (4-20mA input)